SediGraph

View on micromeritics.com ↗

 

Particle size distribution of inorganic particles by x-ray sedimentation (X-ray sedimentation) – an industry standard for more than four decades

 

  • Measurement range 0.1 – 300 µm (equivalent spherical diameter) – Stokes’ settling law with direct mass measurement by X-ray absorption
  • Complete particle accountability – includes fractions outside the direct measurement range
  • Large sample (50 mL of dispersed sample) – representative for heterogeneous materials, no need for precise concentration
  • Optional automation – up to 18 unattended samples with the MasterTech 052 autosampler
  • Compliant with ISO 13317-3, ASTM B761 and ASTM Committee C28 (advanced ceramics)
SediGraph

Details

  • Combining data with other particle size methods extends the reporting range up to 125,000 µm (125 mm) – useful for geological applications
  • Scan of the sedimentation cell from bottom to top – accurate accounting of fast-settling particles with faster resolution of fine particles
  • Fully automatic operation increases sample throughput and reduces the possibility of operator error
  • Temperature-controlled analysis – stable liquid properties during measurement ensure repeatability
  • SPC (statistical process control) reports – monitoring of production processes and fast response to deviations
  • Software: point-and-click interface, customizable reports with a lab logo, data export; graphs – cumulative mass/surface area/number, settling velocity distribution, log-probability, Rosin-Rammler, regression analyses and more; size display also in Phi units
  • Optional MasterTech 052 autosampler – ultrasonic probe for sample redispersion (adjustable power, self-tuning drive circuit), up to 18 samples in sequence, fully unattended
  • SediSperse® – Micromeritics’ line of dispersion liquids (9 aqueous and organic variants) for maximum suspension and prevention of particle agglomeration
  • Typical applications: ceramics, metal powders, geology/soil, cosmetics, pigments, catalysts, construction materials, minerals and inorganic chemicals, abrasives

 

Measurement range 0.1 – 300 µm, equivalent spherical diameter
Sample size 50 mL of dispersed sample – precise concentration not required
Dispersion liquids any liquid compatible with the cell materials and weakly absorptive to X-rays (typically water, glycols, mineral oils, SediSperse®, alcohols)
Wetted materials stainless steel, Teflon-impregnated anodized aluminum, nickel-plated aluminum, nylon, polypropylene, polystyrene, Tygon®/Pharmed® tubing, tungsten carbide, Ertalyte®, Viton®, Buna-n, epoxy
Power 85 – 264 VAC, 47/63 Hz, 450 VA
Temperature operating: ambient +10 to 40 °C · storage/transport: -10 to 55 °C
Humidity 20 – 80% (non-condensing)
Dimensions H 52 × W 50.5 × D 58 cm · weight 43 kg